Product No : TOS2000_Series
  • CE.jpg

Dual channel 20MHz/40MHz/50MHz

High sensitivity 1mV/DIV

10 times sweep magnification

TV synchronization; X-Y mode

Z-axis modulation input; CH1 signal output

High illumination internal graticule CRT

Japanese encoded switch, reliable and durable

Fully sealed durable attenuation switch

ALT triggering function, enabling simultaneous observation of two independent signals

Trigger level lock function; Auto synchronize function

Model

TOS-2020

TOS-2040

TOS-2050

Vertical system

Sensitivity

5mV~5V/DIV, 10 steps in 1-2-5 sequence

Sensitivity accuracy

≤3%; X5MAG: ≤5%

Verniersensitivity

1/2.5 or less than panel indicated value

Bandwidth

DC~20MHz

(X5MAG: DC~7MHz)

DC~40MHz

(X5MAG: DC~15MHz)

DC~50MHz

(X5MAG: DC~15MHz)

AC coupling

 

Rise time

Approx. 17.5ns

(X5MAG: Approx. 50ns)

Approx. 8.75ns

(X5MAG: Approx. 25ns)

Approx. 7ns

(X5MAG: Approx. 23.3ns)

Input impedance

Approx. 1MΩ//Approx. 25pF

DC balance shift

5mV~5V/DIV: ±0.5DIV

Vertical mode

CH1, CH2, DUAL (ALT/CHOP), ADD, CH2 INV

Chopping repetition frequency

Approx. 250kHz

Input coupling

AC, GND, DC

Max. Input voltage

400V peak-peak, AC frequency≤1kHz

Common mode rejection ratio

>50:1 at 50kHz sine wave (set CH1 and CH2 at same sensitivity)

CH2 INV BAL

Balanced point variation≤1DIV (referred to graticule center)

Horizontal system

Sweep time base

0.2us~0.5s/DIV, 20 steps in 1-2-5 sequence

Sweep time base accuracy

±3%, X10MAG: ±5% (20ns~50ns/DIV uncalibrated)

Vernier sweep time base

≤1/2.5 of panel indicated value

Sweep magnification

X10 (Max. sweep time base 20ns/DIV)

Position shift caused by X10MAG

Within 2DIV at CRT screen center

Linearity

±5%, X10MAG: ±10% (0.2s~1us)

Triggersystem

Trigger mode

AUTO, NORM, TV-V, TV-H

Trigger level lock

Provided

Trigger source

CH1, CH2, LINE, EXT

Trigger coupling

AC: 20Hz to full bandwidth

Polarity

“+” or “-” polarity

Sensitivity

Frequency

20Hz~2MHz

2MHz~20MHz

20MHz~40MHz

40MHz~50MHz

CH1, CH2

1DIV

1.5DIV

2.5DIV

3DIV

ALT

2DIV

3DIV

 

 

EXT

200mV

800mV

 

 

TV: Sync pulse>1DIV (EXT: 1V)

External trigger input

Input impedance: Approx. 1MΩ//Approx. 25pF

Max.input voltage: 400V (DC+AC peak), AC frequency≤1kHz

X-Y mode

Sensitivity

5mV~5V/DIV, ±3%

X-axis bandwidth

DC~500kHz (-3dB)

X-Y phase error

≤3° at DC~50kHz

Output signal

CH1 signal

Min.20mV/DIV (50Ω output bandwidth 50Hz~5MHz)

Calibration signal

Waveform

Positive-going square wave

Frequency

Approx. 1kHz

Duty ratio

<48:52

Output voltage

2Vpp±2%

Output impedance

Approx. 1kΩ

Z-axis input

Sensitivity

5Vpp

Bandwidth

DC~2MHz

Input impedance

Approx. 47kΩ

Max. input voltage

30V (DC+AC peak), AC frequency≤1kHz

CRT

Display

6" rectangle, internal graticule, 8x10DIV (1DIV=1cm)

Phosphor

P31

Accelerating voltage

Approx. 2kV (20MHz); Approx. 12kV (40MHz)

Trace rotation

Adjustable at front panel

General

Power source

AC110V/220V±10% selectable, 50/60Hz, Max.35VA

Accessories

Power cord x1, Operation manual x1, Probe x2

Dimension

310x150x455mm

Weight

Approx.8kg

Accessory List

Analog Oscilloscope

Model

Bandwidth

Channel

Special features

TOS-2020

20MHz

2

 

TOS-2040

40MHz

2

 

TOS-2050

50MHz

2

 

Download Details

Catalog

Twintex_catalog_TOS2000